h. At the Analog Sweep menu, press the

menu soft-key

Sweep Ramp

to go to

the Analog Sweep Ramp Menu.

i. Press

Sweep Time

to open the sweep

time parameter for editing.

j. Set the sweep time to 99 seconds.

NOTE

Monitor the 681XXC’s Horizontal
Output on the Oscilloscope to de-
termine sweep start and stop.

Step 4

During the analog sweep, measure the
maximum and minimum Power Meter
readings and record the values on the test
record. Verify that the variation (differ-
ence between the maximum and mini-
mum readings) does not exceed the value
noted on the test record.

5-18

681XXC OM

OPERATION

POWER LEVEL ACCURACY

VERIFICATION

AND FLATNESS TESTS