Erasing
Stored Setups

The front panel setups are stored in non-volatile
memory. A master reset is required to erase the con-
tents of the setups and reprogram them with default
data.

To perform a master reset, proceed as follows:

Step 1

With the 681XXC in standby, press and
hold the RF OUTPUT ON/OFF key.

Step 2

Press the LINE OPERATE/STANDBY
key to turn the instrument on.

Step 3

When the first menu is displayed (after
the start-up display), release the RF
OUTPUT ON/OFF key.

The contents of non-volatile memory have now been
erased and reprogrammed with default data.

3-90

681XXC OM

NOTE

The master reset function over-
writes all information stored in the
non-volatile memory with default
values. This includes the table of
2000 frequency/power level sets
used for the list sweep mode and
the five power-offset tables used for
the user level flatness correction
function.