Step 3

Set up the 681XXC as follows for an ana-
log sweep power level flatness test:

a. Reset the instrument by pressing

SYSTEM

, then

Reset

. The CW

Menu is displayed.

b. Press

Analog

to place the 681XXC in

the analog sweep frequency mode and
display the Analog Sweep Menu.

c. With the Analog Sweep menu dis-

played, press the main menu key

FREQUENCY

CONTROL

The Sweep Frequency Control Menu
(below) is displayed.

d. Press

Full

to select a full range fre-

quency sweep.

e. Press

Edit L1

to open the current

power level parameter for editing.

f. Set L1 to the power level noted on the

test record.

g. Now, return to the Analog Sweep

menu by pressing the main menu key

CW/SWEEP

SELECT

681XXC OM

5-17

OPERATION

POWER LEVEL ACCURACY

VERIFICATION

AND FLATNESS TESTS