TROUBLESHOOTING

Table 6-4 provides procedures for troubleshooting common malfunc-
tions encountered during operation of the signal generator. Included
are procedures for troubleshooting faults that do not produce error
messages, such as, failure to power up and unexpected shutdown.

6-10

681XXC OM

Signal Generator will not turn on

(

OPERATE

light is

OFF

)

Normal Operation: When the 681XXC is connected to the power
source, the

OPERATE

light should illuminate and the instrument

should power up.

Step 1

Disconnect the 681XXC from the power source, then check
the line fuse on the rear panel.

q

If the fuse is defective, replace (see page 6-14).

q

If the fuse is good, go to the next step.

Step 2

Check to see if power is available at the power receptacle.

q

If not, move to a working receptacle.

q

If power is available, go to the next step.

Step 3

Check the power cable.

q

If defective, replace.

q

If good, call a service technician.

Signal Generator will not turn on

(

OPERATE

light is

ON

)

Normal Operation: When the 681XXC is connected to the power
source, the

OPERATE

light should illuminate and the instrument

should power up.

q

If the

OPERATE

light illuminates but the unit fails to

power up, the 681XXC has an internal component fail-
ure. Call a service technician.

Table 6-4.

Troubleshooting (1 of 3)

OPERATOR
MAINTENANCE

TROUBLESHOOTING