Attenuator decoupling 3-64, Attenuator decoupling, 3-64, Attenuator decoupling -64 – Anritsu 681XXC User Manual

Page 97: Local (front leveling panel) operation operations

Advertising
background image

Attenuator
Decoupling

In 681XXCs equipped with option 2 step attenua-
tors, the ALC and attenuator work in conjunction to
provide leveled output power down to –140 dBm. In
the normal (coupled) leveling mode, when the de-
sired power level is set, the correct combination of
ALC level and attenuator setting is determined by
the instrument firmware. In some applications, such
as receiver sensitivity testing, it is desireable to con-
trol the ALC level and attenuator setting separately
by decoupling the step attenuator from the ALC.
The ALC mode menu lets you select attenuator
decoupling.

At the ALC Mode menu, press

Atten Menu

. The

Attenuator Menu (below) is displayed.

This menu lets you decouple the step attenuator
from the ALC, set the power level, and set the at-
tenuation in 10 dB steps.

Press

Decouple

to decouple the step attenuator

from the ALC.

Press

Edit L1

to open the power level parameter for

editing. Edit the current level using the cursor con-
trol key or rotary data knob or enter a new value us-
ing the key pad and appropriate termination key.
When you have finished setting the power level,
press

Edit L1

to close the open parameter.

To change the attenuation setting, press

Incr Atten

or

Decr Attn

. Pressing these soft-keys changes the

attenuation in 10 dB steps.

Press

Previous Menu

to return to the ALC Mode

Menu display.

3-64

681XXC OM

NOTE

The set power level may not be
maintained when switching be-
tween attenuator coupling modes.

LOCAL (FRONT

LEVELING

PANEL) OPERATION

OPERATIONS