681XXC OM

IEEE-488 GPIB

13 14 15 16 17 18 19 20 21 22 23 24

1

2

3

4

5

6

7

8

9 10 11 12

PIN

SIGNAL NAME

SIGNAL DESCRIPTION

1-4

DIO 1 thru DIO 4

Data Input/Output: Bits are HIGH when the data is logical 0 and LOW when the
data is logical 1.

5

EOI

End or Identify: A low-true state indicates that the last byte of a multibyte mes-
sage has been placed on the line.

6

DAV

Data Valid: A low-true state indicates that the active talker has (1) sensed that
NRFD is high-false and NDAC is low-true, (2) placed the data byte on the bus,
and (3) waited an appropriate length of time for the data to settle.

7

NRFD

Not Ready For Data: A high-false state indicates that all active listeners are ready
to accept new data.

8

NDAC

Not Data Accepted: A low-true state indicates that all addressed listeners have
accepted the current data byte for internal processing.

9

IFC

Interface Clear: A low-true state places all bus instruments in a known, quiescent
state—unaddressed to talk, unaddressed to listen, and service request idle.

10

SRQ

Service Request: A low-true state indicates that a bus instrument desires the im-
mediate attention of the controller.

11

ATN

Attention: A low-true state indicates that the bus is in the command mode (data
lines are carrying bus commands). A high-false state indicates that the bus is in
the data mode (data lines are carrying device-dependent instructions or data).

12

Shield

Chassis Ground

13-16

DIO5 thru DIO6

Data Input/Output: Bits are HIGH when the data is logical 0 and LOW when the
data is logical 1.

17

REN

Remote Enable: A low-true state enables bus instruments to be operated re-
motely, when addressed.

18-24

GND

Logic Ground