background image

Press

Slope On/Off

to activate the ALC power slope

function.

Press

Edit Pivot

to open the pivot point frequency

parameter for editing. Edit the current frequency
using the cursor control key or rotary data knob or
enter a new value using the keypad and appropriate
termination key. When you have finished setting the
open parameter, close it by pressing

Edit Pivot

again or by making another menu selection.

Press

Edit Slope

to open the slope parameter for

editing. Edit the current slope value using the cur-
sor control key or rotary data knob or enter a new
value using the key pad and the

STEPS

termination

key. When you have finished setting the open pa-
rameter, close it by pressing

Edit Slope

again or by

making another menu selection.

While monitoring the power level at the device-
under-test (DUT), adjust the power slope and pivot
point to level the power at the DUT.