The 681XXC provides square wave (pulse) modula-
tion of the output signal using modulating signals
from either its internal square wave generator or an
external signal generator.

The signal generator’s internal square wave genera-
tor outputs modulating signals of 400 Hz, 1 kHz,
7.8125 kHz, and 27.8 kHz. The modulating signals
are selectable from a menu.

The 681XXC accepts modulating signals from an ex-
ternal signal generator that are TTL-compatible
with the minimum pulse width of >5

ms.

Providing
Square Wave
Modulation

The following are the menu selections necessary to
provide square wave (pulse) modulation of the out-
put signal using a modulating signal from both the
internal and external sources.

Press

MODULATION

. At the resulting menu dis-

play, press

SqWave

. The Square Wave Status

Menu (below) is displayed.

This menu contains the square wave status window
that shows the current menu selections. This menu
lets you perform the following:

q

Turn square wave modulation on/off.

q

Select Internal or External source for the
modulation signal.

q

Select the polarity of the signal (High or Low)
that turns the RF on.

q

Go to an additional menu (to select the fre-
quency from the internal source or to select the
front or rear panel input connector).

681XXC OM