2 master-slave operation 7-4, Connecting the instruments 7-4, Master-slave operation, 7-4 – Anritsu 681XXC User Manual

Page 208: Master-slave operation -4, Connecting the instruments -4, Use with master-slave other instruments operation

Advertising
background image

7-2

MASTER-SLAVE
OPERATION

Master-slave operation consists of connecting any two 68XXXC and/or
68XXXB instruments together and configuring them so that they pro-
duce CW and synchronized, swept output signals at an operator-
selectable frequency offset. One instrument (the Master) controls the
other (the Slave) via interface cables between their rear panel

AUX I/O

and

SERIAL I/O

connectors. The two units are phase-locked together

by connecting them to the same 10 MHz reference time base.

Connecting
the Instru-
ments

Connect the two instruments, shown in Figure 7-1,
as follows:

Step 1

Connect the 3-port AUX I/O cable end la-
beled “MASTER” to the rear panel

AUX

I/O

connector on the Master instrument.

Connect the AUX I/O cable labeled
“SLAVE” to the rear panel

AUX I/O

con-

nector on the Slave instrument.

Step 2

Connect the ends of the flat interface ca-
ble to the rear panel

Serial I/O

connectors

on the Master and Slave instruments.

Step 3

Connect one end of a coaxial cable to the
rear panel

10 MHz REF OUT

connector on

the Master instrument. Connect the other
end to the rear panel

10 MHz REF IN

con-

nector on the Slave instrument.

7-4

681XXC OM

USE WITH

MASTER-SLAVE

OTHER INSTRUMENTS

OPERATION

Figure 7-1.

68XXXB/68XXXC Configuration for Master-Slave Operation

NOTES

When connecting two instruments
together for Master-Slave opera-
tions, always use a Anritsu
Master-Slave interface cable set,
Part No. ND36329.

If a Model 56100A Scalar Network
Analyzer is being used with the
master-slave configuration, (1) con-
nect the AUX I/O cable end labeled
“SNA” to the rear panel

AUX I/O

connector on the 56100A SNA and
(2) connect a dedicated system bus
cable (P/N 2100-1) between the
Master instrument rear panel

IEEE-488 GPIB

connector and the

56100A SNA rear panel

DEDI-

CATED GPIB

connector.