Before going to the Leveling Menu from the ALC
Mode menu, select whether the external ALC signal
is to be connected to the front- or rear-panel

EXT

ALC IN

connector.

At the ALC Mode menu, press

Ext ALC Front

to se-

lect front panel input, or

Ext ALC Rear

to select

rear panel input.

Now, press

Leveling Menu

to go to the Leveling

Menu (below).

Next, select the type of external sensor you are us-
ing to detect the output power.

To select the external ALC input from an external
detector, press

External Detector

.

To select the external ALC input from a power me-
ter, press

Power Meter

.

After you have made the external ALC input con-
nection and selected the sensor type, press

ALC Loop

. The ALC Loop Menu (below) is dis-

played.

While monitoring the power level at the external de-
tection point, first press

Ext ALC Adj

, then use the

cursor control key or rotary data knob to adjust the
external ALC signal to obtain the set power level.

To return to the Leveling Menu, press

ALC Mode

then press

Leveling Menu

.