Power Level
Flatness Test
Procedure

Power level flatness is checked by measuring the
power level variation during a full band sweep; first
in the step sweep mode, then in the analog sweep
mode.

Step 1

Set up the 681XXC as follows for a step
sweep power level flatness test:

a. Reset the instrument by pressing

SYSTEM

, then

Reset

. The CW

Menu is displayed.

b. Press

Step

to place the 681XXC in

the step sweep frequency mode and
display the Step Sweep Menu (below).

c. With the Step Sweep menu displayed,

press the main menu key

FREQUENCY

CONTROL

The Sweep Frequency Control Menu
(below) is displayed.

d. Press

Full

to select a full range fre-

quency sweep.

e. Press

Edit L1

to open the current

power level parameter for editing.

f. Set L1 to the power level noted on the

test record.

681XXC OM

5-15

OPERATION

POWER LEVEL ACCURACY

VERIFICATION

AND FLATNESS TESTS