Test procedure 5-6, Test procedure, 5-6, Test procedure -6 – Anritsu 681XXC User Manual

Page 150: Operation cw frequency verification accuracy test

Advertising
background image

Test
Procedure

The following procedure tests both the coarse and
fine loops to verify the accuracy of the CW frequency
output.

Step 1

Set up the 681XXC as follows:

a. Reset the instrument by pressing

SYSTEM

, then

RESET

. Upon reset,

the CW Menu (below) is displayed.

b. Press

Edit F1

to open the current fre-

quency parameter for editing.

c. Set F1 to the first test frequency for

the model being tested (Table 5-2A is
the test record for standard models;
Table 5-2B for models with Option 11).

Step 2

Verify that the Frequency Counter read-
ing meets specifications (

±100 Hz of the

value shown on the test record for stan-
dard models;

±10 Hz for instruments with

Option 11).

Step 3

Record the Frequency Counter reading on
the test record (Table 5-2A or Table 5-2B).

NOTE

The Frequency Counter reading is
typically within

±1 Hz. Differences

of a few Hertz can be caused by
noise or counter limitations. Differ-
ences of

³±100 Hz (³±10 Hz for in-

struments with Option 11) indicate
a frequency synthesis problem.

Step 4

Set F1 to the next test frequency on the
test record and record the Frequency
Counter reading.

Step 5

Repeat step 4 until all frequencies listed
on the test record have been recorded.

5-6

681XXC OM

OPERATION

CW FREQUENCY

VERIFICATION

ACCURACY TEST