Place the signal generator in a fixed power level
mode by pressing the main menu key

LEVEL/ALC

SELECT

At the resulting menu display, press

Level

. The

681XXC is now in fixed (non-swept) power level
mode.

At the Level Menu, press

User Cal

. The User Level

Cal Menu (below) is displayed.

This menu lets you perform the following:

q

Create a power-offset table.

q

Select a measurement frequency range (edit
the start and stop frequency parameters).

q

Select the number of points at which correction
information is to be taken.

q

Apply a power-offset table to the test setup.

First, press the menu soft-key to select the power-
offset table (

User 1

,

User 2

,

User 3

,

User 4

, or

User 5

)

that you wish to create.

Next, set the measurement frequency range by
pressing

Edit Start

or

Edit Stop

to open the start

(

Flo

) or stop (

Fhi

) frequency parameter for editing.

Edit the current frequency using the cursor control
key or rotary data knob or enter a new value using
the keypad and appropriate termination key. When
you have finished setting the open parameter, close
it by pressing its menu edit soft-key or by making
another menu selection.

Then, select the number of frequency points at
which correction information is to be taken by press-
ing

Edit Points

to open the number-of-points

parameter for editing. Edit the current number-of
-points using the cursor control key or rotary data
knob or enter a new value using the keypad and the

STEPS

termination key. (The number of points

3-70

681XXC OM

LOCAL (FRONT

LEVELING

PANEL) OPERATION

OPERATIONS