Local (front leveling panel) operation operations – Anritsu 681XXC User Manual

Page 105

Advertising
background image

Erasing the Power-Offset Tables from Memory
The power-offset tables are stored in non-volatile
memory. A master reset is required to erase the con-
tents of the tables and reprogram them with default
data.

To perform a master reset, proceed as follows:

Step 1

With the 681XXC in standby, press and
hold the

RF OUTPUT ON/OFF

key.

Step 2

Press the

LINE OPERATE/STANDBY

key

to turn the instrument on.

Step 3

When the first menu is displayed (after
the start-up display), release the

RF OUT-

PUT ON/OFF

key.

The contents of non-volatile memory have now been
erased and reprogrammed with default data.

3-72

681XXC OM

LOCAL (FRONT

LEVELING

PANEL) OPERATION

OPERATIONS

NOTE

The master reset function over-
writes all information stored in the
non-volatile memory with default
values. This includes the nine
stored front panel setups and the
table of 2000 frequency/power level
sets used for list sweep mode.