g. Now, return to the Step Sweep menu

by pressing the main menu key

CW/SWEEP

SELECT

h. At the Step Sweep menu, press

Sweep Ramp

to go to the Step Sweep

Ramp Menu (below).

i. Press

Dwell Time

to open the dwell

time-per-step parameter for editing.

j. Set the dwell time to 1 second.

NOTE

Monitor the 681XXC’s Horizontal
Output on the Oscilloscope to de-
termine sweep start and stop.

Step 2

As the 681XXC steps through the full fre-
quency range, measure the maximum and
minimum Power Meter readings and rec-
ord the values on the test record. Verify
that the variation (difference between the
maximum and minimum readings) does
not exceed the value noted on the test rec-
ord.

5-16

681XXC OM

OPERATION

POWER LEVEL ACCURACY

VERIFICATION

AND FLATNESS TESTS