Power level accuracy test procedure 5-14, Accuracy test procedure, 5-14, Power level accuracy test procedure -14 – Anritsu 681XXC User Manual

Page 158

Advertising
background image

Power Level
Accuracy Test
Procedure

Power level accuracy is checked by stepping the
power down in 1 dB increments from its maximum
rated power level.

Step 1

Set up the 681XXC as follows:

a. Reset the instrument by pressing

SYSTEM

, then

Reset

. The CW

Menu (below) is displayed.

b. Press

Edit F1

to open the current fre-

quency parameter for editing.

c. Set F1 to the CW frequency noted on

the test record for the model being
tested.

d. Press

Edit L1

to open the current

power level parameter for editing.

e. Set L1 to the power level noted on the

test record.

Step 2

Measure the output power level with the
Power Meter and record the reading on
the test record.

Step 3

Verify that the Power Meter reading
meets the specifications stated on the test
record.

Step 4

Set L1 to the next test power level. Re-
cord the Power Meter reading on the test
record.

Step 5

Repeat step 4 for the other levels listed on
the test record for the current CW fre-
quency.

Step 6

Repeat steps 1 thru 5 for all CW frequen-
cies listed on the test record.

5-14

681XXC OM

OPERATION

POWER LEVEL ACCURACY

VERIFICATION

AND FLATNESS TESTS

NOTE

In models with Option 22 that have
a high-end frequency of

£20 GHz,

rated output power is reduced by
1 dB. In models with Option 22 that
have a high-end frequency of
>20 GHz, rated output power is re-
duced by 2 dB.