681XXC OM

6-11

Signal Generator Quits During Operation

(

OPERATE

light remains on)

Trouble Description: The signal generator operates for some time,
then shuts down (

OPERATE

light remains on). After a short period, the

signal generator resumes normal operation. This is an indication that
the 681XXC has reached an excessive operating temperature.

Step 1

Check that the fan is still operating during the time that the
instrument is shut down.

q

If the fan is still operating, clean the air filter (see page
6-13).

q

If the fan is not operating, call a service technician.

LOCK ERROR

is Displayed

Trouble Description: This message is displayed in the frequency pa-
rameters area to indicate that the output frequency is not phase-
locked. It is normally caused by an internal component failure.

Step 1

Perform a self-test of the signal generator by pressing the
System Menu soft-key

Selftest

.

q

If self-test does not result in an error message(s), resume
normal operation.

q

If an error message(s) is displayed, call a service techni-
cian.

Table 6-4.

Troubleshooting (2 of 3)

OPERATOR
MAINTENANCE

TROUBLESHOOTING