Local (front signal panel) operation modulation – Anritsu 681XXC User Manual

Page 109

Advertising
background image

This menu contains an external FM status window
that shows the current menu selections and the
measured FM Deviation. (The FM deviation meas-
urement function measures the voltage of the exter-
nal modulation signal and calculates the peak
frequency deviation.) This menu lets you perform
the following:

q

Turn FM on/off.

q

Set the FM sensitivity.

q

Select the input impedance (600

W or 50W) of

the input connector.

q

Select the input connector (front panel or rear
panel

FM IN

) that is connected to the external

signal source.

Press

On / Off

to turn FM on and off. Both the FM

status display and FM modulation status area will
reflect your selection.

Press

Edit Sens

to open the FM Sensitivity pa-

rameter, then edit the current value using the cur-
sor control key or rotary data knob or enter a new
value using the keypad and the appropriate termi-
nator key. The FM Sensitivity range is

±10 kHz/V to

±20 MHz/V for Locked and Unlocked Narrow FM
modes and

±100 kHz/V to ±100 MHz/V for Unlocked

Wide FM mode. To close the open FM Sensitivity pa-
rameter, press

Edit Sens

or make another menu se-

lection.

Press

600

W/50W to select the input impedance of

the input connector. The FM status display will
reflect your selection.

Press

Front/Rear

to select the front or rear panel

FM IN

connector. The FM status display will reflect

your selection.

3-76

681XXC OM

LOCAL (FRONT

SIGNAL

PANEL) OPERATION

MODULATION

ERR

This error message is displayed
when the external FM modulating
signal exceeds the input voltage
range. The message “

Reduce FM

Input Level

” also appears at the bot-

tom of the FM status display. FM is
turned off until the modulating sig-
nal is within the input voltage
range.