3-11

LEVELING
OPERATIONS

The 681XXC generates leveled output power over a maximum range of
up to 33 dB (up to 149 dB with Option 2). Instruments with Option
15A provide leveled output power over a maximum range of up to
27 dB (up to 141 dB with Option 2). An automatic level control (ALC)
system controls the amplitude and power level of the RF output. The
operator can select the ALC mode of operation—internal, external
(detector or power meter), or fixed gain (ALC off). In addition, the sig-
nal generator provides (1) an ALC power slope function that provides
compensation for high frequency system or cable losses, (2) a decouple
function that allows decoupling of the step attenuator (if equipped)
from the ALC system, and (3) a user level (flatness correction) calibra-
tion function that provides for calibrating out path variations with fre-
quency in a test setup.

The following paragraphs provide descriptions and operating instruc-
tions for the power leveling modes and functions. Use the Leveling
Modes menu map (Chapter 4, Figure 4-10) to follow the menu se-
quences.

Selecting a
Leveling
Mode

The ALC system is a feedback control system, in
which the output power is measured at a detector
and compared with the expected power level. If the
output and desired power levels do not equal, the
ALC adjusts the power output until they do. The
feedback signal can be provided by either the inter-
nal detector or an external detector or power meter.
Alternatively, the output power can be set to a fixed
level without using the normal feedback (ALC off).
The ALC mode menu lets you make the selection of
a leveling mode.

To go to the ALC Mode menu, first press the main
menu key

LEVEL/ALC

SELECT

At the Level/ALC Select Menu display, press

ALC Mode

. The ALC Mode Menu (below) is

displayed.

3-60

681XXC OM

LOCAL (FRONT

LEVELING

PANEL) OPERATION

OPERATIONS