16 reference oscillator calibration 3-92, Calibration, reference oscillator, 3-92, Reference oscillator calibration, 3-92 – Anritsu 681XXC User Manual

Page 125: Reference oscillator calibration -92

Advertising
background image

3-16

REFERENCE
OSCILLATOR
CALIBRATION

The reference oscillator calibration function lets you calibrate the
internal 100 MHz crystal reference oscillator of the 681XXC using an
external 10 MHz, 0 to +10 dBm reference signal.

NOTE

Before beginning calibration, always let
the 681XXC warm up for a minimum of
120 hours.

To perform calibration of the internal reference os-
cillator, first connect the external 10 MHz reference
signal to the 681XXC rear panel

10 MHz REF IN

con-

nector.

Next, press the

SYSTEM

main menu key. At the

System Menu display, press

Cal Menu

to go to the

Calibration Menu (below).

Press

Refrnce Cal

to begin calibration.

Press

Previous Menu

to return to the System Menu

display.

When

Refrnce Cal

is pressed, the Calibration

Status Menu (below) is displayed.

Press

Proceed

to start the calibration.

Press

Abort

to cancel the calibration and return to

the Calibration Menu display.

3-92

681XXC OM

LOCAL (FRONT

REFERENCE

PANEL) OPERATION

OSCILLATOR CALIBRATION

NOTE

This calibration is not applicable to
units having Option16, High Sta-
bility Time Base, installed.