Performance specifications, B-8 681xxc om – Anritsu 681XXC User Manual

Page 237

Advertising
background image

Warm Up Time:

From Standby:

30 minutes.

From Cold Start (0

°

C):

120 hours to achieve specified

frequency stabiltiy with aging.
Instruments disconnected from ac line power for more
than 72 hours require 30 days to return to specified fre-
quency stability withaging.

Power:

90-132 Vac or 180-264 Vac, 48–440 Hz, 400 VA maximum

Standby: With ac line power connected, unit is placed in
standby when front panel power switch is released from the
OPERATE position.
Weight: 23 kg maximum
Dimensions:

133 H x 429 W x 597 D mm

RF Output Connector:

Type K female,

£

40 GHz models

Type V female, >40 GHz models.

ENVIRONMENTAL

Storage Temperature Range: –40

°

C to +75

°

C.

Operating Temperature Range: 0

°

C to +50

°

C.

Relative Humidity: 5% to 95% at 40

°

C.

Altitude: 4,600 meters.

EMI

Meets the radiated emission requirements of:

EN55011:1991/CISPR-11:1990 Group 1 Class A
EN50082-1:1997/

EN 61000-4-2:1995 - 4 kV CD, 8 kV AD
EN 61000-4-3:1997 - 3 V/m
ENV 50204 - 3 V/m
EN 61000-4-4:1995 - 0.5 kV SL, 1 kV PL
EN 61000-4-5:1995 - 1 kV L-L, 2 kV L-E

MIL-STD-461C Part 2 RE01, RE02, CE01, CE03,
CS01, CS02, CS06, RS03

INPUTS and OUTPUTS

AM IN : Accepts an external signal to AM modulate the RF
output signal. Front or rear-panel input, 50

W

or 600

W

impedance, both selectable from front-panel modulation
menu.
FM IN : Accepts an external signal to FM modulate the RF
output signal. Front or rear-panel input, 50

W

or 600

W

impedance, both selectable from front-panel modulation
menu.

IN : Accepts an external TTL compatible signal to

pulse modulate the RF output signal. Front or rear-panel in-
put, selectable from front-panel modulation menu.
EXT ALC IN (External ALC Input): Provides for leveling
the RF output signal externally with either a detector or
power meter. Signal requirements are shown in the RF Out-
put specifications on page B-5.
RF OUTPUT: Provides for RF output from 50

W

source

impedance. K or V Connector, female. Option 9 moves the
RF Output connector to the rear panel.
10 MHz REF IN: Accepts an external 10 MHz

±

100 Hz,

0 to +10 dBm time-base signal. Automatically disconnects
the internal high-stability time-base option, if installed. 50

W

impedance.
10 MHz REF OUT: Provides a 0.5 Vp-p, AC coupled,
10 MHz signal derived from the internal frequency standard.
50

W

impedance.

HORIZ OUT (Horizontal Sweep Output): Provides 0V at
beginning and +10V at end of sweep for all sweep modes,
regardless of sweep width. In CW mode, the voltage is pro-
portional to frequency between 0V at low end and +10V at
the high end of range. In CW mode, if CW RAMP is en-
abled, a repetitive, 0V to +10V ramp is provided.

B-8

681XXC OM

PERFORMANCE
SPECIFICATIONS

Input/Output Connectors

Nomenclature

Type

Location

AM IN

BNC

Front & Rear Panel

FM IN

BNC

Front & Rear Panel

IN

BNC

Front & Rear Panel

EXT ALC IN

BNC

Front & Rear Panel

RF OUTPUT

K-Connector
V-Connector

Standard-Front Panel

Option 9-Rear Panel

10 MHz REF IN

BNC

Rear Panel

10 MHz REF OUT

BNC

Rear Panel

HORIZ OUT

BNC

Rear Panel

AUX I/O

25-pin D-type

Rear Panel

SERIAL I/O

RJ45

Rear Panel

IEEE-488 GPIB

Type 57

Rear Panel

Advertising