5-5

CW FREQUENCY
ACCURACY TEST

The following test verifies that the CW frequency output of the signal
generator is within accuracy specifications. Table 5-2, beginning on
page 5-7, contains test records that you can copy and use to record test
results for this test. Test records for standard 681XXC models are con-
tained in Table 5-2A ; test records for 681XXC models with Option 11
are contained in Table 5-2B.

Test Setup

Connect the equipment, shown in Figure 5-1, as fol-
lows:

Step 1

Connect the 681XXC rear panel

10 MHz

REF OUT

to the Frequency Counter 10

MHz External Reference input. If the
Frequency Counter has an INT/EXT tog-
gle switch, ensure the switch is set to
EXT.

Step 2

Connect the 681XXC

RF OUTPUT

to the

Frequency Counter RF Input as follows:

a. For measuring frequencies of 0.01 to

1.0 GHz, connect to the Band 2 input
(Connection A).

b. For measuring frequencies of 1.0 to

26.5 GHz, connect to the Band 3 input
(Connection A).

c. For measuring frequencies of 26.5 to

65.0 GHz, connect to the Band 4 input
as shown in Connection B using the
appropriate waveguide mixer; Option
91 (26.5 to 40 GHz), Option 92 (40 to
60 GHz), or Option 93 (60 to 90 GHz).

681XXC OM

5-5

OPERATION

CW FREQUENCY

VERIFICATION

ACCURACY TEST

6 8 1 X X C S I G N A L G E N E R A T O R

F R E Q U E N C Y C O U N T E R

R F I N

B a n d 2

I n p u t

B a n d 3

I n p u t

C o n n e c t i o n A

( 0 . 0 1 t o 2 6 . 5 G H z )

C o n n e c t i o n B

( 2 6 . 5 t o 6 5 G H z )

1 0 M H z

E X T I N

B a n d 4

I n p u t

M i x e r

1 0 M H z

R E F O U T

R F O U T

Figure 5-1.

Equipment Setup for CW Frequency Accuracy Test