Local (front leveling panel) operation operations – Anritsu 681XXC User Manual

Page 104

Advertising
background image

range is 2 to 801.) When you have finished setting
the open number-of-points parameter, close it by
pressing

Edit Points

or by making another menu

selection.

Now, press

Start Cal

to begin automatically taking

power level correction information at each frequency
point. During this process the menu displays the
status:

Calibrating

along with the current measure-

ment frequency point.

NOTE

To terminate the measurement process at
any time before completion, press

Abort

.

Once the power-offset table has been created, it is
stored in non-volatile memory. The power-offset ta-
ble is now ready to be applied to the test setup. Dis-
connect the Power Sensor and Power Meter from the
test setup.

Applying User Level Flatness Correction
Whenever user level flatness correction is applied to
the test setup by activating the power-offset table,
the set power level is delivered at the point where
the calibration was performed.

To activate the selected power-offset table and apply
user level flatness correction to the test setup, press

On/Off

. The User Level Cal menu will display the

status:

On.

To turn off the selected power-offset table and re-
move user level flatness correction from the test
setup, press

On/Off

again. The User Level Cal

menu will display the status:

Off

.

Entering a Power-Offset Table via GPIB
User level flatness correction can be applied to the
test setup using a power-offset table created from
calculated data and entered via the GPIB. Refer to
the 681XXC Programming Manual (P/N 10370-
10334) for information and instructions on creating
a power-offset table and entering it via the GPIB.

681XXC OM

3-71

LOCAL (FRONT

LEVELING

PANEL) OPERATION

OPERATIONS

USER 1...5

When a power-offset table is se-
lected ON, this status message is
displayed on all menu displays to
remind the operator that user level
flatness correction has been applied
to the ALC.