6 power level accuracy and flatness tests 5-13, Test setup 5-13, Test setup, 5-13 – Anritsu 681XXC User Manual

Page 157: Power level accuracy and flatness, 5-13, Power level accuracy and flatness, Tests -13, Test setup -13

Advertising
background image

5-6

POWER LEVEL
ACCURACY AND
FLATNESS TESTS

These tests verify that the power level accuracy and flatness of the
681XXC meet specifications. Table 5-3, beginning on page 5-19, con-
tains test records that you can copy and use to record test results for
these tests. Test records are provided for each 681XXC model configu-
ration.

Test Setup

Connect the equipment, shown in Figure 5-2, as
follows:

Step 1

Calibrate the Power Meter with the
Power Sensor.

NOTE

For

£40 GHz models, use the

M A 2 4 7 4 A p o w e r s e n s o r ; f o r
>40 GHz models, use the MA2475A
power sensor.

Step 2

Connect the Power Sensor to the

RF OUT-

PUT

of the 681XXC.

Step 3

Connect the 681XXC rear panel

HORIZ

OUT

to the Oscilloscope CH.1 input (X in-

put).

NOTE

Before starting these procedures,
locate the test record in Table 5-3
for the particular 681XXC model
configuration being tested.

681XXC OM

5-13

OPERATION

POWER LEVEL ACCURACY

VERIFICATION

AND FLATNESS TESTS

R F

O U T

H O R I Z

O U T

6 8 1 X X C S I G N A L G E N E R A T O R

O S C I L L O S C O P E

P o w e r

S e n s o r

P O W E R M E T E R

C H 1 o r X I n p u t

Figure 5-2.

Equipment Setup for Power Level Accuracy and Flatness Tests