Setting up the 681xxc 7-17, Setting up the 681xxc -17 – Anritsu 681XXC User Manual

Page 221

Advertising
background image

Step 3

Connect the special AUX I/O interface
cable (Anritsu Part No. 806-90) to the
681XXC rear panel

AUX I/O

connector.

Connect the cable end having BNC con-
nectors as follows:

a. Connect the cable end labeled “SEQ

SYNC” to the HP8757D rear panel

POS Z BLANK

connector.

b. Connect the cable end labled “SWP

DWELL IN” to the HP8757D rear
panel

STOP SWEEP

connector.

c. Connect the cable end labled “HORIZ

OUT” to the HP8757D rear panel

SWEEP IN 0 - 10V

connector.

Setting Up
the 681XXC

The 681XXC must be set to GPIB address 19 and in
the 8757D Scalar mode of operation to operate as a
signal source for the SNA. The following paragraphs
describe how to set up the 681XXC to enable the
8757D Scalar GPIB mode.

On the 681XXC front panel, press

LINE

to place the

signal generator in operation.

Allow the signal generator to warm up, then press
the

SYSTEM

main menu key. At the System Menu

display, press

Config

. The System Configuration

Menu (below) is displayed.

At the System Configuration menu, press

GPIB

.

The Configure GPIB Menu (on the following page) is
displayed.

681XXC OM

7-17

USE WITH

USE WITH A HP8757D

OTHER INSTRUMENTS

SCALAR NETWORK ANALYZER