Setting power level step size 3-59, Setting step size, 3-59, Setting power level step size -59 – Anritsu 681XXC User Manual

Page 92

Advertising
background image

Setting Power
Level Step
Size

There are two ways to set the step size of the power
level step that occurs after each frequency sweep
—set the step size or set the number of steps. The
step size range is 0.01 dB (Log) or 0.001 mV (Linear)
to the full power range of the ssignal generator; the
number of steps range is 1 to 10,000. The power
level step size is set from the level sweep ramp
menu.

To go to the Level Sweep Ramp Menu (below) from
the Level Sweep menu, press

Sweep Ramp

.

This menu lets you set the step size and the number
of steps.

Press

Step Size

to open the step size parameter.

Press

Num of Steps

to open the number of steps pa-

rameter.

Open the parameter you wish to change, then edit
the current value using the cursor control key or ro-
tary data knob or enter a new value using the key-
pad and appropriate termination key. When you
have finished setting the open parameter, close it by
pressing its menu soft-key or by making another
menu selection.

Press

Previous Menu

to return to the Level Sweep

Menu display.

681XXC OM

3-59

LOCAL (FRONT

POWER LEVEL

PANEL) OPERATION

SWEEP OPERATION

RANGE

This error message is displayed
when (1) the step size value entered
is greater than the level sweep
range or (2) the number of steps en-
tered results in a step size of less
than 0.01 dB (Log) or 0.001 mV
(Linear). Entering a valid step size
will clear the error.