Performance specifications, B-6 681xxc om – Anritsu 681XXC User Manual

Page 235

Advertising
background image

User Level Flatness Correction:

Number of points: 2 to 801 points per table
Number of tables: 5 available
Entry modes: GPIB power meter or computed data

CW POWER SWEEP
Range:
Sweeps between any two power levels at a single
CW frequency.
Resolution: 0.01 dB/step (Log) or 0.001 mV (Linear)
Accuracy: Same as CW power accuracy.
Log/Linear Sweep: Power sweep selectable as either log
or linear. Log sweep is in dB; linear sweep is in mV.
Step Size: User-controlled, 0.01 dB (Log) or 0.001 mV
(Linear) to the full power range of the instrument.
Step Dwell Time: Variable from 1 ms to 99 seconds. If the
sweep crosses a step attenuator setting, there will be a
sweep dwell of approximately 20 ms to allow setting of the
step attenuator.

SWEEP FREQUENCY/STEP POWER
A power level step occurs after each frequency sweep.
Power level remains constant for the length of time re-
quired to complete each sweep.

MODULATION

AMPLITUDE MODULATION
All amplitude modulation specifications apply at 50%
depth, 1 kHz rate, with RF level set 6 dB below maximum
specified leveled output power, unless otherwise noted.
External AM Input: Log AM or Linear AM input, front or
rear-panel BNC, 50

W

or 600

W

input impedance. All options

selectable from modulation menu.
AM Sensitivity:

Log AM:

Continuously variable from 0 dB/volt to

25 dB/volt
Linear AM:

Continuously variable from 0% per volt to

100% per volt

AM Depth (typical): 0-90% linear; 20 dB log
AM Bandwidth (3 dB): DC to 50 kHz minimum

DC to 100 kHz typical

Flatness (DC to 10 kHz rates):

±

0.3 dB

Accuracy:

±

5%

Distortion: <5% typical
Incidential Phase Modulation

(30% depth, 10 kHz rate): <0.2 radians

Maximum Input:

±

1V

FREQUENCY MODULATION
External FM Input:
Front or rear panel BNC, 50

W

or 600

W

input impedance. All options selectable from modulation
menu.
FM Sensitivity: Continuously variable from

±

10 kHz per

volt to

±

20 MHz per volt (Locked or Unlocked Narrow FM

modes) or

±

100 kHz per volt to

±

100 MHz per volt (Un-

locked Wide FM mode), selectable from modulation menu.
Maximum FM Deviation:

Locked Mode (1 kHz to 500 kHz rates):
The lesser of

±

10 MHz or rate x 300

Unlocked Narrow Mode (DC to 500 kHz rates):

±

10 MHz

Unlocked Wide Mode (DC to 100 Hz rates):

±

100 MHz

FM Bandwidth (3 dB):

Locked Mode:

1 kHz to 500 kHz

Unlocked Narrow Mode:

DC to 500 kHz

Unlocked Wide Mode:

DC to 100 Hz

Flatness (10 kHz to 500 kHz rates):

±

1 dB

Accuracy (100 kHz rate,

±

1V input): 10% (5% typical)

Maximum Input:

±

1V

SQUARE WAVE MODULATION
The RF output can be pulse modulated via an external
modulating signal or an internal square wave generator.
On/Off Ratio:

>50 dB

Rise/Fall Time: <1

m

s typical

Internal Square Wave Generator: Four square wave sig-
nals (400 Hz, 1 kHz, 7.8125 kHz, and 27.8 kHz), selectable
from modulation menu.

Accuracy:

Same as internal or external 10 MHz time

base.
Square Wave Symmetry:

50%

±

5% at all power levels

External Input: Front or rear-panel BNC, selectable from
modulation menu.

Drive Level:

TTL compatible input

Minimum Pulse Width:

>5

m

s.

Input Logic:

Positive-true or negative-true, selectable

from modulation menu.

B-6

681XXC OM

PERFORMANCE
SPECIFICATIONS

Advertising