VOLTAGE MEASUREMENT (RCD, LOOP, PHASE ROTATION)

Range [V]

Resolution [V]

Accuracy

0

÷ 460V

1

±(3%Reading + 2digit)

LOOP P-P, P-N: LINE IMPEDANCE MEASUREMENT (Phase – Phase, Phase - Neutral)

Range [

] Resolution

[

]*

Accuracy

0.01

ч 9.99

10.0

ч 199.9

0.01

0.1

±(5% Reading + 3digit)

(*) 0.1 m

Ω on range 0.0 ÷ 199.9 mΩ (with IMP57)

Peak value of the Test current

127V

3.65A

230V

6.64A

400V

11.5A

Voltage Range (Phase – Phase, Phase – Neutral)

100

ч250/100ч440V

Frequency 50Hz

± 0.5Hz

LOOP P-PE: FAULT LOOP IMPEDANCE MEASUREMENT (Phase - Earth)

Range [

] Resolution

[

]*

Accuracy

0.01

ч 19.99

20.0

ч 199.9

200

ч 1999

0.01

0.1

1

±(5% Reading + 3digit)

(*) 0.1 m

Ω on range 0.0 ÷ 199.9 mΩ (with IMP57)

Peak value of the Test current:

127V

3.65A

230V 6.64A

Voltage Range (Phase –Earth)

100

÷250V

Frequency 50Hz

± 0.5Hz

LOOP R

a 15mA

: FAULT LOOP IMPEDANCE MEASUREMENT without RCD Tripping (Phase - Earth)

Range [

] Resolution

[

]

Accuracy

1

÷ 1999

1

- 0%, +(5% Reading + 3digit)

Test Current

15mA

Voltage Range (Phase –Earth)

100

÷250V 50Hz

EARTH: GROUND RESISTANCE MEASUREMENT WITH EARTH RODS

Range RE [

]

Resolution [

]

Accuracy

0.01 – 19.99

20.0 – 199.9

200 - 1999

0.01

0.1

1

±(5% Reading + 3 digit)

Test Current

<10mA – 77.5Hz

Open circuit Test Voltage

<20V RMS

RESISTIVITY MEASUREMENT

Range

ρ

Resolution

Accuracy

0.60

÷19.99 Ωm

20.0

÷ 199.9Ωm

200

÷ 1999Ωm

2.00

÷ 99.99kΩm

100.0

÷ 125.6kΩm(*)

0.01

Ωm

0.1

Ωm

1

Ωm

0.01 k

Ωm

0.1 k

Ωm

±(5% Reading + 3 digit)

(*) setting distance = 10m
Test Current

<10mA – 77.5Hz

Open circuit Test Voltage

<20V RMS


14.1.2. ANALYZER and AUX functions

VOLTAGE MEASUREMENT – SINGLE PHASE SYSTEM (AUTORANGE)

Range [V]

Resolution [V]

Accuracy

Input Impedance

15

÷ 310V

0.2V

310

÷ 600V

0.4V

±(0.5%

Reading+2digit)

300k

Ω (Phase-Neutral)

300k

Ω (Phase-Phase)

VOLTAGE SAG AND SURGE DETECTION – SINGLE PHASE SYSTEM (MANUAL RANGE)

Range [V]

Resolution

(Voltage)

Resolution

(Time)

Accuracy

(Voltage)

Accuracy (rif. 50Hz)

(Time)

Input Impedance

15

÷ 310V

0.2V

30

÷ 600V

0.4V

10ms (½ period)

±(1.0% rdg +2 digit)

± 10ms (½ period)

300k

Ω (Phase-Neutral)

300k

Ω (Phase-Phase)

CURRENT MEASUREMENT with FlexEXT and STD clamps – SINGLE PHASE SYSTEM (AUTORANGE)

Range [V] (*)

Resolution [mV]

Accuracy

Input Impedance

Overload Protection

0.005

÷0.26V 0.1

0.26

÷1V 0.4

±(0.5%rdg +2digit)

400k

5V

(*): Example: with a 1000A/1V full scale clamp , the instrument detect only current higher than 5A

CURRENT MEASUREMENT with FlexINT clamps - 1000A range

Range (A)

Voltage input

Resolution

Accuracy

10.0

÷19.9

950.0

µV ÷ 1.691mV

±(4.0% rdg + 8.5µV)

20.0

÷ 99.9

1.7mV

÷ 8.491mV

±(1.0% rdg + 8.5µV)

100.0

÷ 999.9

8.5mV

÷ 84.99mV

8.5

µV

±(1.0% rdg + 8.5µV)

1A = 85

µV ; Rinput = 400kΩ