HT instruments SIRIUS89N User Manual

Page 36

Advertising
background image

SIRIUS89N

EN - 34

RCD 05.06.01



- - -

ms


FRQ=50.0Hz Ut= ---V
VP-N=234V VP-PE= 234V

o UT DANGEROUS

x1 30mA

50V

!

The instrument doesn't
detect an efficient
protection circuit.

FUNC

IdN RCD UL

If a contact voltage

U

t

higher than the
selected limit (U

L

) is

detected, the instrument

interrupts the test and
emits

a long sound

signal at the end of the
test and displays the
screen alongside.

RCD 05.06.01



- - -

ms


FRQ =50.0Hz Ut= ---V
VP-N=234V VP-PE= 34V

o NO PE

x1 30mA

50V

!

Message

“NO PE”: the

instrument doesn't detect
an efficient protection
circuit.

FUNC

IdN RCD UL

If the instrument detects
that the earth cable
(green) is not
connected, the screen
alongside is displayed
for 5 seconds then the
initial display is back.
Check the connections
of PE conductor under
test.

RCD 05.06.01



- - -

ms


FRQ=50.0Hz Ut= ---V
VP-N=231V VP-PE= 230V

o HOT

x1 500mA

50V

!

FUNC

IdN RCD UL

Message “hot”: the
instrument got overheated.

If the instrument gets
overheated, tests cannot
be carried out and the
message alongside is
displayed. Wait until the
initial screen is back in
order to proceed with
measurements.

THE PREVIOUS RESULTS CANNOT BE SAVED.