17. APPENDIX 1 – MESSAGES DISPLAYED

Message Description

Advices

AUTONOM:

Available memory autonomy for the recording which is

being effected

CLEAR ALL? (Enter)

The operator is trying to cancel all the recordings effected

Press ESC in order not to cancel the whole

memory, press ENTER to confirm

CLEAR LAST? (Enter)

The operator is trying to cancel the last recording effected

Press ESC in order not to cancel the last

recording, press ENTER to confirm

Data saved

The data have been saved

DATA SIZE:

Dimensions of the stored data

HOLD

By pressing the proper key, the HOLD function has been

activated

Press HOLD again to disable this function

Password:

A recording has been started and at least 5 minutes have

passed from the last activity of the instrument (see

paragraph 7).

Insert the password:

F1, F4, F3, F2

Invalid date

The inserted date is not correct

Check the inserted date

Energy Measuring

The instrument is taking an energy measurement

Press F1 to stop it

Memory Full

The memory of the instrument is full

Cancel some recordings after transferring
them to a PC

No ext supply!

A recording has been started without connecting the

external power supply (optional code A0050)

Verify it you really want to start the recording

without the external power supply. In that
case press START again.

No parameter sel

A recording has been started without selecting any value to

be recorded

Press START/STOP and select at least a

value entering the MENU

No Phase selected

Voltage and/or current harmonics have been selected and

the corresponding flag has been enabled (HARMONICS

ON) but no phase voltage or current has been selected

Select at least one phase voltage and/or

current

PASSWORD ERROR

The inserted password is wrong (see paragraph 7).

Check the password

PASSWORD OK

The inserted password is correct

Please wait

The instrument is waiting for the recording to be started
(see paragraph 6)

Recording

The instrument is recording (see paragraph 6)

Too many param

More than 63 parameters have been selected (harmonics

included) or More than 38 parameters with CO-

GENERATION Flag enabled

Deselect some values

Too many records

The quantity of recorded data + Smp exceeds the

maximum allowed (35)

Cancel some recordings after transferring

them to a PC

No Unit selected

ERR: SEQ

The Phase Sequence isn’t correct.

Check the Phase Sequence connection.

ERR: P-

The active powers shown on the right side of the message
are negative

If there isn’t a situation of co-generation
check if the clamps are properly connected

ERR: SEQ & P-

The active powers shown on the right side of the message

are negative and the Phase Sequence isn’t correct.

If there isn’t a situation of co-generation

check if the clamps are properly connected /
check the Phase Sequence connection.

ERR: CONNECTION

The instrument has detected a wrong connection to

Voltage inputs

Check the Voltage connections

Error Vref

The user set a Voltage reference not compatible with

voltage at instrument’s input.

Check Voltage Reference set in “CONFIG

RECORDER”

ERR: SYNC

The System Frequency is out of range

Check the System Frequency, check setting

in ANALYZER CONFIG.

Selection Error

There is a mismatch between the Parameter enabled and

the parameter selected for an AUX recording.

Check the parameter enabled in AUX

position and the selected parameter for

recording.

Error1

÷ Error 5

The instrument memory is damaged.

Contact HT Italia assistance