8.8.2. Using the Instrument in a Three Phase System

CAUTION

The maximum voltage between B1 and B4 inputs is 600 V~ (CATII) / 350V~
phase – earth or 600V~ (CATIII) / 300 V~ phase to earth.
Do not measure voltages exceeding the limits prescribed by this manual.
Should you exceed the voltage limits you could damage the instrument
and/or its components or endanger your safety.


The instrument is a single phase system. So you can analyse a three phase system
repeating 3 times a single phase analysis described in previous paragraph.

Instrument connection in a three-phase system


The Three Phase power values are given by:

Total Active Power = Pt = P1 + P2 + P3

Total Reactive Power = Qt= Q1 + Q2 + Q3

Total Apparent Power = St =

(Pt

2

+ Qt

2

)