HT instruments SIRIUS89N User Manual

Page 64

Advertising
background image

SIRIUS89N

EN - 62

5. Press

START/STOP to stop the measurement and display the message

"

waiting for final SPL" and then the following screen will be displayed:

05.06.01 11:43:04

Sound Pressure Level

Peak Value (dBA)

Measurement Duration

SOUND


Leq 94.2dB

Peak 121.7dB


Duration 0000:00:00

MEASURING...

PG+

.


It

contains:

- Duration: duration of the measurement expressed in hours : minutes : seconds
- L

Eq

: value of equivalent noise level during the measurement.

[

]



=

+ 0

1

1

2

10

0

/

)

(

0

1

log

10

t

t

t

eq

dt

p

t

p

t

L

- Peak: peak value of SPL.

6. The displayed result can be stored pressing the

SAVE key twice (according to

paragraph 9.1).


7.3.1.1. SOUND LEVEL MEASUREMENT "faulty cases"

SOUND 05.06.01


SPL ----dB

Peak ----dB

Duration 0000:00:00

o

HT55:NO RS232

!

PG*

If the instrument
communicate with the
probe, the message
"HT55: NO RS232" will
be displayed.

The instrument can't
communicate with Probe.
Check if the Batteries aren't
weak, the cable and the
adapter are correctly
connected.

THE PREVIOUS RESULTS CAN'T BE SAVED.