6.1.3. Results of "AUTO" mode

LOW

Ω 05.06.01

Average resistance value
(

Ravg)

Resistance values and
corresponding test current
got exchanging the
polarities of test leads



1.05Ω


R+ R-
1.07

Ω 1.03Ω

219mA 219mA


AUTO 0.11

FUNC

CAL

At the end of the test, if
the

average resistance

value Ravg is lower
than 5

Ω the instrument

emits a

double sound

signal indicating the
positive outcome of
the test
and displays
one screen similar to the
screen alongside.

The displayed result can be stored pressing the

SAVE key twice (according to

paragraph 9.1).


6.1.4. Results of "RT+" and "RT-" modes

LOW

Ω 05.06.01

Max Resistance value of
R+ or R-.

Test current

Duration of the Test



1.07Ω

219mA






RT+ 0.11

Ω TIME: 10s

FUNC

CAL

if a

resistance value

RT+ or RT+ lower than
5

is detected, the

instrument emits a
double sound signal
indicating the positive
outcome
of the test
displays one screen
similar to the screen
alongside.


Note:

We recommend to use crocodiles and to check if the crocodiles make a good
contact with the conductor under test. Indeed in this test the instrument gives as
final result the maximum measured value of R+ or R- and suing test leads
instead of crocodiles could give you faulty result due to faulty contact between
the test leads and conductor under test

The displayed result can be stored pressing the

SAVE key twice (according to

paragraph 9.1).