Sirius89n, En - 102 – HT instruments SIRIUS89N User Manual

Page 104

Advertising
background image

SIRIUS89N

EN - 102

CURRENT MEASUREMENT with FlexINT clamps - 3000A range

Range (A)

Voltage input

Resolution

Accuracy

30.0

÷ 999.9

2.55mV

÷ 84.99mV

8.5

µV

±(1.0%rdg + 17µV)

1000

÷ 3000

85.0mV

÷ 255mV

85

µV

±(0.5% rdg + 85µV)

1A = 85

µV ; Rinput = 400kΩ

POWER MEASUREMENT – SINGLE PHASE SYSTEM (AUTORANGE)

Quantity Range

Accuracy

MAX

Resolution

ACTIVE POWER

0

÷ 999.9W

1

÷ 999.9kW

1

÷ 999.9MW

1000

÷ 9999MW

0.1W

0.1kW

0.1MW

1MW

REACTIVE POWER

0

÷ 999.9VAR

1

÷ 999.9kVAR

1

÷ 999.9MVAR

1000

÷ 9999MVAR

0.1VAR

0.1kVAR

0.1MVAR

1MVAR

APPARENT POWER

0

÷ 999.9VA,

1

÷ 999.9kVA,

1

÷ 999.9MVA

1000

÷ 9999MVA

0.1VA

0.1kVA

0.1MVA

1MVA

ACTIVE ENERGY
(Classe2 EN61036)

0

÷ 999.9Wh,

1

÷ 999.9kWh,

1

÷ 999.9MWh

1000

÷ 9999MWh

0.1Wh

0.1kWh

0.1MWh

1MWh

REACTIVE ENERGY
(Classe3 IEC1268)

0

÷ 999.9VARh,

1

÷ 999.9kVARh,

1

÷ 999.9MVARh

1000

÷ 9999MVARh

±(1.0%Reading+2digit)

0.1VARh

0.1kVARh

0.1MVARh

1MVARh

Cos

ϕ MEASUREMENT – SINGLE PHASE SYSTEM

Cos

ϕ

Resolution Accuracy

[°]

0.20

ч 0.50

1.0

0.50

ч 0.80

0.7

0.80

ч 1.00

0.01

0.6

VOLTAGE AND CURRENT HARMONICS MEASUREMENT – SINGLE PHASE SYSTEM

Range

Accuracy Resolution

DC – 25H

±(5% rdg + 2 digit)

26H – 33H

±(10% rdg + 2 digit)

34H – 49H

±(15% rdg + 2 digit)

0.1V / 0.1A

Harmonics values are null under fixed threshold:

- DC: its values is null if it is < 2%of Fundamental or is <2% of Full Scale clamp

- 1

st

Current Harmonic: its values is null if it is < 0.2% Full Scale clamp

- 2

nd

÷ 49

th

: its values is null if it is < 2% of fundamental or is <2% of Full Scale clamp

ENVIRONMENTAL PARAMETER MEASUREMENT

Range

Accuracy

Resolution

-20°C –80 °C

0.1 °C

0

÷ 100% UR

0.1% UR

0.001Lux

÷ 20.00 Lux

0.001

÷ 0.02 Lux

0.1Lux

÷ 2000 Lux

0.1

÷ 2 Lux

1Lux

÷ 20 kLux

±(2% rdg + 2dgt)

1

÷ 20 Lux

LEAKAGE CURRENT MEASUREMENT

Range (*)

Resolution [mA]

Accuracy

Input Impedance

Overload Protection

0.5

÷ 999.9mA

0.1mA

±(5% rdg + 2digit)

400k

Ω 5V

(*): During a recording the instrument detect only Current > 5mA with Resolution 1mA