O > 42 – HT instruments SIRIUS89N User Manual

Page 38

Advertising
background image

SIRIUS89N

EN - 36

RCD 05.06.01


RCD tripping time is bigger
than the maximum
measurable time (it
depends on type of test,
see following table).



o

>999

ms


FRQ=50.0Hz Ut= 1V
VP-N=231V VP-PE= 230V

o

TIME NOT OK

x1 30mA

50V

!

FUNC

IdN RCD UL

ATTENTION: the tripping
time is higher than the
standard limit.

If the RCD tripping time
is higher than the
instrument’s measuring
limits, the instrument
emits

a long sound

signal at the end of the
test and displays the
values alongside.


The maximum duration depends on the test type:

Test type

General RCD

Selective RCD

MAN x1 test

999ms

999ms

MAN x2 test

200ms

250ms

MAN x5 test

50ms

160ms

" " test

300ms

RCD 05.06.01

Tripping Current.


27

mA

>300ms


Freq=50.0Hz Ut= 1V
VP-N=231V Vp-PE=230V

o TIME NOT OK

30mA

50V

!

FUNC IdN RCD

UL

Tripping Time exceeding
limit value

During the ramp test
if the RCD tripping time
is higher than the limit,
the instrument emits

a

long sound signal at
the end of the test and
displays the values
alongside.


RCD 05.06.01

Maximum current
generated by the
instrument during the test
for general RCDs (the
value indicated is referred
to an AC type 30mA RCD,
in this case the maximum
current supplied is equal to
1.4xI

∆N


o > 42

mA

>300ms


FRQ=50.0Hz Ut= 1V
VP-N=231V VP-PE=230V

o CURRENT NOT OK

30mA

50V

!

FUNC

IdN RCD UL

During the ramp test
if the RCD tripping
current is higher than I

∆n

(Type AC) or 1.4 I

∆n

(Type A with I

∆n

>10mA)

or 2 I

∆n

(Type A with

I

∆n

≤10mA) , the

instrument emits

a long

sound signal at the end
of the test and displays
the values alongside.

ATTENTION:
the RCD tripping current is
higher than the Nominal

Value (I

∆N

=30mA was set

in the example

).

The previous results can be stored pressing the

SAVE key twice (according to

paragraph 9.1).

Advertising