HT instruments SIRIUS89N User Manual

Page 23

Advertising
background image

SIRIUS89N

EN - 21

6.2.2. Results of "MAN" mode

M

Ω 05.06.01

Insulation Resistance

Voltage during the Test

Duration of the Test

Test mode



1.07

M



514V 15s


MAN 500V

Test voltage set

FUNC

VNOM

At the end of the test if
the Insulation
resistance is

lower

than R

MAX

(see Table2)

and the instrument
generated the
Nominal test Voltage
,
the instrument emits a
double sound signal
indicating the positive
outcome of the test

and displays one
screen similar to the
screen alongside.


In order to evaluate the test you must compare the result with the limits indicated in the
Guidelines (see Table1).

The displayed result can be stored pressing the

SAVE key twice (according to

paragraph 9.1).

M

Ω 05.06.01

Maximum resistance value
which can be measured
(999

Ω is displayed if a

rated voltage of 500V was
selected see Table2).

The symbol

">" means that

the resistance value is
higher than R

MAX

.



> 999

M




523V 15s


MAN 500V

Test duration

FUNC

VNOM

If the Insulation
resistance is

higher

than R

MAX

(see Table2),

the instrument emits a
double sound signal at
the end of the test
indicating the positive
outcome of the test

and displays one screen
similar to the screen
alongside.

The displayed result can be stored pressing the

SAVE key twice (according to

paragraph 9.1).