6.2.4. "MAN" and "TIMER" mode faulty cases

M

Ω 05.06.01

Disconnect the External Power
Supply Adapter

If the instrument
detect the External
Power supply adapter
connected to
instrument will show
the message
displayed to side.




-.- -

M




---V 15s

o REMOVE POWER


MAN 500V

!

FUNC VNOM

M

Ω 05.06.01

ATTENTION: the test can't be
executed. Check that the circuit is
not energised.



-.- -

M




---V 15s

o VOLT IN INPUT


MAN 500V

FUNC VNOM

If the instrument
detect a Voltage
between Input
terminals higher than
15V, the instrument
does not effect the
test and displays the
screen alongside for 5
seconds.

These result can't be saved


M

Ω 05.06.01

Insulation Resistance



o 1.17

M




107V 15s



MAN 500V

ATTENTION: the test of
resistance

R

ISO

was effected at

a voltage value lower than the
set rated voltage
. Low
insulation case. This case
occurs under low insulation
conditions or in the presence of
capacity on the installation.

FUNC VNOM

If the instrument can't
generate the Nominal
Test Voltage it will
emits a

long

acoustic signal and
displays a screen
similar to the screen
alongside.

Test Time

The displayed result can be stored pressing the

SAVE key twice (according to

paragraph 9.1).