Caution – HT instruments SIRIUS89N User Manual

Page 17

Advertising
background image

SIRIUS89N

EN - 15

6.1.1.1. Procedure to reset test leads calibration parameters

LOW

Ω 05.06.01

Message

>99.9

Ω:

means that the
instrument detected a
resistance higher
than 5

Ω therefore it

will proceed with
Reset procedure.



o

>

99.9


R+ R-
----

Ω ----Ω

---mA ---mA


AUTO 0.11

FUNC

CAL

To cancel calibration
parameters it is
necessary to effect a
calibration procedure
with a resistance of
test leads higher than
5

Ω (for example with

open test leads). When
a cancellation is effected
the screen nearby is
displayed first

, followed

by the screen below:

6.1.2. Measurement Procedure

1. Select the desired mode by means of the

F1 key.

2. Connect the black and blue test leads to

B1 and B4 input terminals respectively

Connection of the test leads during LOW

test.

3. If the cables supplied with the instrument are not long enough for the

measurement you can extend the blue cable.

4. Short-circuit the test leads making sure that the conductive parts of the

crocodiles make a good contact to each other. Press the START key.

If the

displays doesn't show 0.00

repeat the test leads calibration (see

paragraph 6.1.1).

5. Connect the instrument terminals to the ends of the conductor under test (see

previous picture).

6.

If the mode "RT+" or "RT-" was selected use the F3, F4 keys to set the
duration of the test.

7. Press the

START key. The instrument will execute the measurement. In

RT+/RT- (Timer mode) you can press START key again if you want to stop the
test before the duration set is expired.

CAUTION

Never disconnect the test leads when the message

"MEASURING" is

displayed.

START

STOP