HT instruments SIRIUS89N User Manual

Page 46

Advertising
background image

SIRIUS89N

EN - 44

6.4.7. LOOP

Faulty Cases

LOOP 05.06.01

Disconnect the External
Power Supply Adapter

If the instrument detect
the External Power
supply adapter
connected to instrument
will show the message
displayed to side.

- - -

---A



FRQ=50.0HZ
VP-N=231V VP-PE=230V

o REMOVE POWER


P-N

!

FUNC Z2


LOOP 05.06.01

NO VOLTAGE Detected

Should the instrument
detect that the phase
and/or neutral cables
are not connected to an
installation, screen
alongside is displayed
when pressing START.


- - -

---A



FRQ=50.0HZ
VP-N= 1V VP-PE= 0V

NO VOLTAGE


P-PE 50V

!

FUNC Z2

UL


LOOP 05.06.01

HIGH VOLTAGE Detected


- - -

---A



Freq=50.0HZ
VP-N=401V VP-PE=230V

HIGH VOLTAGE


P-PE 50V

!

FUNC Z2

UL

Should the instrument
detect a voltage
between phase and
neutral higher of 250V,
for example in case the
blue cable is connected
to an installation phase
conductor of a 400V
three-phase system, the
screen alongside is
displayed.