Recordings, Caution – HT instruments SIRIUS89N User Manual

Page 91

Advertising
background image

SIRIUS89N

EN - 89

10. RECORDINGS

10.1. START A RECORDING

The recording function is available for

ANALYZER and AUX rotary switch position.

As you can read in the paragraphs 7.2.1 and 8.2, a recording can be started manually or
automatically. Therefore, after setting all the parameters

and leaving the Menu, the

instrument will start to record:

MANUALLY:

the recording will start when Instrument time reach the “00” seconds

value after pressing

START/STOP.

AUTOMATICALLY:

If the operator has pressed START/STOP the instrument will

remain in stand-by until the date and time previously set, then the recording will start.
While

if the operator doesn’t press START/STOP the recording will never start.


CAUTION

For recordings

we recommend to use the external power supply

adapter (optional code A0050) even the instrument allows the operator to
perform a recording using internal batteries.


If you press Start a recording without the external power supply adapter (optional code
A0050) the instrument will display a warning message “

No ext supply”. Press START key

again to run the recording or press

ESC to quit.



If during a recording the external power supply adapter (optional code A0050) is de-
energised, the instrument will continue the recording using the internal battery power until
the batteries are exhausted (the data stored until the definitive turning off won’t get lost).
For this we recommend you

ALWAYS insert a new set of batteries before a long

recording.

The instrument uses sophisticated algorithms to prolong the battery life. Particularly:

The instrument switches OFF the backlight Automatically after 5 seconds.
If the Battery level is too low the Backlight function will be disabled.
If the instrument is just displaying in real time (and the external power supply is not

connected), after about 5 minutes from the last key pressure or switch rotation the
instrument turns off automatically ("AUTOPOWER OFF" function).

If the instrument is recording or is measuring energy (and the external power supply is

not connected), after about 5 minutes from the last key pressure or switch rotation the
instrument starts a special procedure to save the batteries ("ECONOMY MODE"): the
instrument keeps recording but the display is turned off.


Before starting a recording the operator should first evaluate the state of the equipment,
decide what to record and set the instrument accordingly.