6 temperature control unit x, Temperature control unit x – Eppendorf epBlue with MultiCon User Manual

Page 123

Advertising
background image

epBlue settings - Checking components and making the basic settings

epBlue™ with MultiCon

English (EN)

123

11.6

Temperature control unit X

The

Temperature control unit X

function allows you to check the epMotion thermal modules.

Abb. 11-5: Temperature control unit function

Fig. 11-5:

Temperature control unit

function

The current status will be displayed in the

State

area.

You can make settings in the

Set Temperature

area.

1. Enter the set temperature in the

New Temperature

field.

2. Enter the test duration in the

New time

field.

3. Click on the

Start

button. The thermal module heats or cools until the set temperature has been reached

or the set time has elapsed. If the set temperature is reached before the set time has elapsed, the
thermal module maintains the set temperature until the end of the test.

4. To update the display during the test, press the

Refresh

button.

5. To activate the fan, press the

Fan On

button.

Desired Temperature

Shows the set temperature of the thermal
module.

Actual Temperature

Shows the set temperature of the thermal
module.

Actual current

Shows the current strength in the thermal
module.

Time

Shows the temperature control duration in min.

Remain time

Shows the time remaining until the end of
temperature control (in min).

Temp reached after

Shows the time required to reach the set
temperature.

New Temperature

Set temperature (–5°C to 110°C)

New time

Temperature control duration (0 to 99 min)