6.3.13

Vacuum

command

The

Vacuum

command is used to create a vacuum in the vacuum unit. You can aspirate liquid in filter

plates.

Prior to executing the application, equip the vacuum unit as described in the epMotion hardware operating
manual.

Abb. 6-22: Vacuum window

Fig. 6-22:

Vacuum

window

The command

Vacuum

only is available for devices with vacuum unit.

Frame

drop-down list

Type of Vac Frame used.

Pressure

field

Vacuum in the vacuum unit. Unit mbar or kPa.
Maximum underpressure 850

mbar/85

kPa.

Duration

field

Time during which the vacuum is maintained.
0

s – 100

min

59

s

range.

Use Vacuum Lid

checkbox

Activate transport of the Vac Lid with the gripper.

Use Vacuum Lid

drop-down list

Vac Lid used.

Check levels

checkbox

The optical sensor checks the remaining volume
of all wells in a filter plate on the vacuum unit. If
remaining liquid has been detected, a vacuum
will be reapplied. Checking the remaining volume
takes too long.