0x2135

Location for
destination not
defined

Edit mode/parameter in
command

Transport

:

The destination location for
the labware to be transported
was not edited.



See explanation in
"Cause".

0x2136

Invalid entry for
seconds (1

...

59)

Edit mode/parameter in
command

Wait

:

A value beyond the allowed
range has been entered for the

seconds

parameter.



Enter a value between
0 and 59 seconds.

0x2137

Invalid entry for
minutes (1

...

99)

Edit mode/parameter in
command

Wait

:

A value above the allowable
maximum minutes has been
entered for the

minutes

parameter.



Enter a value between
0 and 99 minutes.

0x2138

Method without
active commands

Application only contains
passive commands (such as
wait, comment,

etc.).



Insert at least one active
command.

0x2139

Parameter conflict:
mix before
aspirating not
allowed in
multiaspirate mode

Pool/Pool One Destination:
Mix before aspirating not
allowed in multiaspirate mode.



Do not mix.

0x2152

Editing vac lid
option in vacuum
command not
possible because
no vac lid on the
worktable

To use the Vac Lid option in
the

vacuum

command the

worktable must be equipped
with a vac lid.



Editing the worktable
please place a Vac Lid on
position "GRIPPER".

0x2154

Parameter conflict:
Check levels time
may not be shorter
than remove lid
time

The

check levels

option can

only be performed when the
lid has been taken off the filter
plate before by the gripper.



Select a

check levels

time

greater than the

remove lid

time.

0x2170

Parameter conflict:
Parameter elution
from filter is only
possible when filter
plates have been
selected as source

To edit the

elution from filter

option a filter plate must have
been edited as source.



See explanations in
"Cause".

Code

Symptom/message Cause

Remedy