10 hepa air filter, 1 switching on the hepa air filter – Eppendorf epBlue with MultiCon User Manual

Page 128

Advertising
background image

epBlue settings - Checking components and making the basic settings
epBlue™ with MultiCon
English (EN)

128

The tool holder moves. The UV lamp will be switched on.

The

UV Light

button shows the remaining irradiation period. The lamp symbol appears in the status area.

11.10

HEPA air filter

Air is drawn in using a fan. This aspirated air is filtered by the HEPA air filter. A continuous and filtered air
flow is guided into the epMotion work room.

11.10.1 Switching on the HEPA air filter

Prerequisites

• The epMotion is equipped with a CleanCap.
• All surfaces in the epMotion work room are disinfected.
• The front hood is closed.
• The epMotion is ready for operation.
• epBlue was started.



Select the

Settings

symbol on the start screen.



On the

epMotion serial number

tab, select the

HEPA filter

function.



Press the

HEPA filter ON

button.

Abb. 11-9: Lamp symbol in status area

Fig. 11-9: Lamp symbol in status area

Abb. 11-10: UV Light button

Fig. 11-10:

UV Light

button

WARNING! Risk of infection due to aerosols.
If you are using a HEPA air filter, a continous air flow is blown from the working chamber of
the epMotion into its surroundings. The air flow can transport aerosols and contaminate the
ambience.



Do not use the HEPA air filter when working with biohazardous substances.



Desinfect contaminated surfaces in the working chamber of the epMotion before switching
on the HEPA air filter.