Eppendorf epBlue with MultiCon User Manual

Page 81

Advertising
background image

epBlue Application Editor - Creating and editing applications

epBlue™ with MultiCon

English (EN)

81

6.5.1.1 Setting application parameters for the

Detect volumes

function

Abb. 6-34: Properties window

Fig. 6-34:

Properties

window

Tab. 6-5:

Detect volumes



To set the application parameters for the

Detect volumes

function (see Defining the application

parameters of the labware on p. 44).

Off

• Deactivate the optical sensor.
• The volume of all vessels must be entered when the application is

started.
For labware with ≤ vessels, the volume for each vessel can be
specified in the

Volume

column.

For labware with > 56 vessels, one volume can be specified for all
vessels. This volume will be recommended at the start of the
application, and can be changed.

• The volume specification does not take into account the surface

meniscus of the liquid.

• If the volume is not enough, the result may be faulty dispensing.

All positions

• Check the volume of all positions in a labware.
• If you work with an eight-channel dispensing tool, the epMotion uses

the calculated volume for each vessel.

• If you work with a multi-channel dispensing tool, the epMotion uses

the lowest calculated volume in a column for the liquid aspiration
and the highest calculated volume in a column for the liquid
dispensing.

If the filling levels in one plate differ significantly or the labware has very low filling levels, use
the

Aspirate from bottom

and

Dispense from top

option. Do not use the

Random access

(see

Sample transfer > Options command on p. 50) option.