12 thermomixer command, Thermomixer command – Eppendorf epBlue with MultiCon User Manual

Page 66

Advertising
background image

epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)

66

6.3.12

Thermomixer

command

The

Thermomixer

command can be used to set the temperature and speed of the thermomixer.

Abb. 6-20: Thermomixer window

Fig. 6-20:

Thermomixer

window

The command

Thermomixer

only is available for devices with thermomixer.

Template

drop-down list

The templates for the labware contain saved
settings for mixing.

Mixing on

checkbox

Activate the mixing function of the thermomixer.

Speed

field

Set the mixing frequency. 300

rpm

2000

rpm

range.

Duration

field

Set the mixing time. 5

s –

120

min

range.

Temperature on

checkbox

Activate the temperature control of the
thermomixer.

Keep temperature after applicaton run

field

Activate the temperature control for the period
after the end of the application. This function can
be used to cool samples or reagents after the end
of the application.

Temperature

field

Set the set temperature of the thermomixer,
4

°C

95

°C

range.

Information on the thermomixer can be found in the epMotion hardware operating manual.

NOTICE! Contamination of samples and the device due to high mixing frequency.
If the speed is set too high, liquid will spray out of the vessels. Labware may become loose
from the holder and fly about.



Observe the maximum speeds of the labware in the operating instructions.