6.3.3

Dilute

command

The

Dilute

command is used to create serial dilution series, e.g., for PCR standards. A defined volume is

pipetted several times from one vessel to the next.

Abb. 6-13: Principle of the Dilute command

Fig. 6-13: Principle of the

Dilute

command

The parameters of the command

Dilute

are identical to the command

Sample transfer

(see p. 48).

If the

Dilute

command is carried out within one labware the source positions and destination positions must

not overlap.

1. Fill the destination positions with a diluent. Use the

Reagent transfer

command.

2. Define the number of samples to be diluted using the

Number of samples

command.

3. In the pattern, define the number of dilution steps. The pattern is limited by a series or column.

4. Enter the options for the

Dilute

command.

5. To achieve a good mixture of sample and diluent, activate the

Mix

>

Mix after dispensing

option.

The sample and diluent will be mixed after each dispensing.

6. To mix the undiluted sample before the first aspiration, activate the

Mix before aspirating

option.

Proceed as follows to enter the pattern for the

Dilute

command:

1. Select the source position of the undiluted sample.

2. Select the destination positions of the dilution steps. The number of positions defines the number of

dilution steps.

3. Repeat the procedure for each source position.

Undiluted Sample

1. Dilution

2. Dilution

3. Dilution

Diluent

Diluent

Diluent