6.3.1.1

Sample transfer

>

Parameter

command

Abb. 6-8: Sample transfer > Parameter

Fig. 6-8:

Sample transfer

>

Parameter

Volume

Select volume in µL or nL. A decimal place is available for volumes ≤

99.9

µL.

You can transfer a volume larger than the maximum volume of the pipette tip.
The volume to be transferred will be pipetted in several steps.
Example: The user would like to transfer 1500

µL

and

enters

this

value

into

the

Volume

field. If a 1000

µL tip is being used, 2x

750

µL will be pipetted. If a

300

µL tip is being used, 5

x

300

µL will be pipetted.

Pipet. tool

Used dispensing tool.

Pipette

The volume is individually aspirated and dispensed for each liquid transfer.

Multidispense

Aspirate liquid and dispense it in the destination position in defined subsets.

Multiaspirate

Aspirate liquid from several source positions using one pipette tip and
dispense the liquid in one destination position. Available for

Pool

and

Pool one

destination

commands.

Use filter tips

Activate when using tips with filter.

Source

Select the source labware. The source labware must be placed on the
worktable. You can select 4 source labwares. If you are using more than one
source labware, the number of positions in the labware (384, 96, 24) must
match. If you dispense within one labware, the source labware and destination
labware are identical.

Destination

Select the destination labware. The destination must be placed on the
worktable. You can select 4 destination labwares. If you are using more than
one destination labware, the number of positions in the labware (384, 96, 24)
must match. If you dispense within one labware, the source labware and
destination labware are identical.

Pattern

Select the source positions and destination positions. Create replicates.

Information on the aspiration and transfer of liquids by means of the epMotion can be found in