8 magnetic separation, Magnetic separation – Eppendorf epBlue with MultiCon User Manual

Page 126

Advertising
background image

epBlue settings - Checking components and making the basic settings
epBlue™ with MultiCon
English (EN)

126

The current status will be displayed in the

State

area.

You can make settings in the

Set Temperature

area.

1. Enter the set temperature in the

New Temperature

field.

2. Enter the mixing time in the

New time

field.

3. Press the

Start

button.

The thermomixer tempers the labware to the set temperature. If the temperature is reached before the
mixing time has elapsed, the thermomixer maintains a constant temperature until the end of the test.

11.8

Magnetic separation

The

Magnetic Separation

function can be used to check the magnetic finger module.

Abb. 11-8: Magnetic Separation function

Fig. 11-8:

Magnetic Separation

function

Home

Move the magnetic finger module to the home
position.

Draw Out

Draw out the magnetic fingers.

Draw In

Draw in the magnetic fingers.