0x211D

Mixing cycles in
source not defined

Parameter

No. of Cycles

in a

mix procedure was not edited
for the source in the
application (mix procedure as
defined in a command

Mix

or

as part of a liquid transfer
command via parameter

Options

).



See explanation in
"Cause".

0x211E

Invalid entry for
mixing cycles in
source (1

...

99)

Entry for the

No. of Cycles

parameter in a mix procedure
for source vessels was higher
than the max. limit (1

up

to

99

cycles) (mix procedure as
defined in a command

Mix

or

as part of a liquid transfer
command via parameter

Options

).



Enter a number between
1 and 99 for the

No. of

Cycles

parameter.

0x211F

Invalid entry for
mixing speed in
source (1

...

10)

Entry for the parameter

Speed

in a mix procedure for source
vessels was higher than the
max. limit (1

up

to

10)

(mix

procedure as defined in a
command

Mix

or as part of a

liquid transfer command via
parameter

Options

).



Enter a number between
1 and 10 for the

Speed

parameter.

0x2120

Mixing volume in
source not defined

Parameter

Volume

in a mix

procedure for source vessels
was not edited in the
application (mix procedure as
defined in a command

Mix

or

as part of a liquid transfer
command via parameter

Options

).



See explanation in
"Cause".

0x2121

Parameter conflict:
mixing volume in
source too large for
this tool

The

Volume

and

Pipet. Tool

parameters of a mix procedure
for source vessels are not in
agreement (

Volume

is higher

than the upper limit of the
tool's volume range) (mix
procedure as defined in a
command

Mix

or as part of a

liquid transfer command via
parameter

Options

).



Edit

Volume

and

Pipet. Tool

so that

Volume

is within

pipette tool's volume
range.

Code

Symptom/message Cause